Spectra Apex HSR Testing
Picture
Picture
Picture
Picture

Spectra Apex HSR Spectral Content
Using multiple Zooming Windows, the Spectral Content of the Spectra Apex HSR is between 20 Hz - 20,000 Hz.
Picture

Spectra Apex HSR Electrosmog Test
The high frequency EMF graphs for the Spectra (Below right) showed very little difference from Background noise (Below Middle). This was confirmed with a Trifield meter on the RF setting which yielded a very small level of .042 mW/m^2.
Picture
Picture
Picture

Spectra Apex HSR Mat MAX Slew Rate and Intensity 
These are the test results from measuring the Spectra Apex HSR PEMF device. The measurements were performed using the AKM EQ730L Quantum Well Hall Element sensor. The signal was read on an Oscilloscope. The sensor is specified to produce 130mV/mT at 5V. This test was performed at 5V, so for every gauss the sensor produces 13mV. The Apex HSR was set to intensity level 100.
Below is the measurement on the surface at the point of maximum intensity (over the windings of the coil).
Spectra Apex HSR Mat MAX Slew Rate and Intensity
​
Below is the measurement on the surface at the point of maximum intensity (over the windings of the coil).
399.62 mV divided by 13mV per gauss equals 30.74 gauss intensity. The pulse rises in these measurements is in the range of 98 us to 104us.

For this measurement, the slew rate is 30.74 T/S.
Picture

Spectra Apex HSR Mat CENTER Slew Rate and Intensity - 0"
Below is the measurement at the surface in the center of the coil:
312.1 mV divided by 13mV per gauss equals 24.0 gauss intensity.
For this measurement, the rise time is 1.0 us, yielding a slew rate of 24.0 T/s.
(2.40 mT/.100 ms = 24.0 T/s)
Picture
Spectra Apex HSR Mat Center Slew Rate and Intensity - 1"
Below is the measurement at the surface in the center of the coil:
270.3 mV divided by 13mV per gauss equals 20.8 gauss intensity.
The pulse rises in these measurements is in the range of 98 us to 104us (or .102 milliseconds ms).

For this measurement, the rise time is 1.01 us, yielding a slew rate of 20.59 T/s.
Picture
Spectra Apex HSR Mat Center Slew Rate and Intensity - 2"
​Below is the measurement at the surface in the center of the coil:
242.7 mV divided by 13mV per gauss equals 18.7 gauss intensity.
For this measurement, the rise time is 1.03 us, yielding a slew rate of 18.15 T/s.
Picture
Spectra Apex HSR Mat Center Slew Rate and Intensity - 3"
​Below is the measurement at the surface in the center of the coil:
199.9 mV divided by 13mV per gauss equals 15.4 gauss intensity.
For this measurement, the rise time is 1.03 us, yielding a slew rate of 14.93 T/s.
Picture
Spectra Apex HSR Mat Center Slew Rate and Intensity - 4"
Below is the measurement at the surface in the center of the coil:
188.6 mV divided by 13mV per gauss equals 14.5 gauss intensity.
For this measurement, the rise time is 1.04 us, yielding a slew rate of 13.95 T/s.
Picture
Spectra Apex HSR Mat Center Slew Rate and Intensity - 5"
​Below is the measurement at the surface in the center of the coil:
157.9 mV divided by 13mV per gauss equals 12.1 gauss intensity.
For this measurement, the rise time is .98 us, yielding a slew rate of 12.32 T/s.
Picture


Spectra Apex HSR Loop  Measurements Coming Soon!
Spectra Apex HSR Loop MAX Surface Slew Rate and Intensity 
Picture
Spectra Apex HSR Loop Surface Slew Rate and Intensity - 0" 
Picture
Spectra Apex HSR Loop Surface Slew Rate and Intensity - 1" 
Picture
Spectra Apex HSR Loop Surface Slew Rate and Intensity - 2" 
Picture
Spectra Apex HSR Loop Surface Slew Rate and Intensity - 3" 
Picture
Spectra Apex HSR Loop Surface Slew Rate and Intensity - 4" 
Picture
Spectra Apex HSR Loop Surface Slew Rate and Intensity - 5" 
Picture